Chapter 40 and 6 rad physics

Why is traditional x-ray not great for vascular imaging?

It doesn't take pictures fasten off, it can't withstand heat, and it's not as flexible.

What are the three types of film for vascular imaging?

Rapid film changers, cut and roll film changers, cineangiographic cameras

What are other names for rapid changers?

Serial changers, or film changers

When was roll film changers developed?

As early as 1925

What parts are included on a roll film changer?

Grid, upper and lower intensifying screens, pressure plate, film cutter, supply and receiving magazines

What is the roll film changers max exposure?

12 per second

What is another name for cut film changers?

Puck film changer

What do you cut film changers allow?

Real-time imaging of angiograms

What is the main difference between roll film changers and cut film changers?

Cut film changers are precut films at the size of 14 x 14. Roll film changers are 14 x 17 that need to be cut.

What are the included parts of a cut some changer?

Grid, two intensifying screens, compression plate, compression table, supply magazines, receiving cassette

Why is it important to have good screen contact?

Because it will be blurry if it is not tight enough

What is the maximum exposure for cut film changers?

Six per second

What is the disadvantage of cut film changers?

They get jammed

What can rapid film changers be used in?

Either monoplane or biplane imaging

What is biplane?

When you have two planes, AP and lateral, taking images at the same time, with one contrast injection

What are the three components of cineangiography imaging systems?

Automatic brightness control, beam splitting mirror, cine camera filming device.

What is automatic brightness control?

It is used to compensate for KVP, mAs, or both, Giving you the best image possible, due to differences in anatomy.

What is the beams splitting mirror?

And allows you to view and record at the same time

What is cine camera filming device?

It takes many single images and move them fast enough, so it looks like one image. Usually at 15, 30, 60 per second. But when you have higher frame rate you have higher patient does.

What is synchronized system?

It means, the x-ray tube is energized only during the time when the cine film is in place for exposure

What needs to be included for digital image acquisition?

You need high frequency generator, x-ray tube, I I, image processor

When angio work is high-powered where shall it deliver?

Anywhere from 800-1000

Why are high frequency/high voltage generator is the most efficient?

Because they never drop to zero

What is the range of KV to better allow better visual of contrast in soft tissue?

50 to 100

What is it called when a generator switches on/off very rapidly to prevent you from overheating?

Pulse progressive fluoroscopy

What is interrogation time?

Time required to prepare x-ray tube to be switched on and reach the selected KVP and MA

What is extinction time?

The time required to stop exposure or shut off x-ray tube

What does three phase 12 poles or multi phase offer?

Constant potential 12 pulses per second and short exposure times at .001 ms

What is short term loading ability?

Amount of heat and anode can tolerate for a sequence of exposures/x-rays

What is continuous heat loading ability?

Amount of heat the anode can tolerate during a fluoro procedure

How big is a large diameter massive anode disc?

11 inches

What is the massive anode disc made of?

Tungsten, molybdenum, and rhenium

Maximum heat loading characteristics of an anode or measured in:

Heat units

What are the three x-ray tube rating charts?

Radiographic rating chart, Ano to cooling chart, housing cooling chart

What is the most important x-ray tube rating chart?

Radiographic rating chart

What do all radiographic tube charts plot?

MA, kVp, and time

What does each filament of each tube have?

A unique radiographic tube chart

What is the formula for heat unit?

HU = kVp x mA x time x rectification factor (x exposures)

What does
Single phase =
3 P 6 P =
3 P 12 P =
High frequency =

1
1.35
1.41
1.45

What does an anode cooling chart tell you?

How long it takes the anode to cool

What does a housing cooling chart tell you?

How long it takes to housing to cool, this is not used today

What does the heat loading capacity tube depend on?

Focal spot size, anode angle, anode speed, and generator

Why must you have a very small focal spot capability?

To view small vessels, to have more spatial resolution, so you get the best detail possible

What must you use if you have an analog image intensifier?

CCD and ADC

What does CCD replace?

TV camera tube

What is ADC?

Analog to digital converter

CCD connects to the ________ which connects________ then that_________ and that __________.

Output phosphor
Light to electrons
Signal is analog
Changes to CCD

What don't digital image intensifiers need?

Conversions, because they will go straight to the digital image

What is the last parts of digital image acquisition system?

Imaging processing hardware and computer

What are the modes of digital image acquisition?

Digital angiography
DSA
Digital Cineradiography

What does DSA stand for and what is it?

Digital Subtraction Angiography- process of removing or masking unnecessary anatomy, and Enhancing only anatomy of interest

What is digital angiography?

When you do one image at a time because it's used when only a few images is needed for placement of a PICC

What does PICC stand for?

Peripherally Inserted Central Catheter

What is mask/scout image?

The initial exposure before and contrast is injected

What is native image(s)?

Images with contrast

cineangiographic camera:

A specialized camera used for cardiac procedures to produce motion picture image of the flow of contrast medium through a vessel

What mm is used with cineangiographic camera?

35 mm roll is commonly used; 16 mm is also used, but not as much; higher dose with 35 mm

Cine Equipment has special tube rating charts and anode cooling charts; what are these known as?

Percent duty factor/cycle

What does window level control?

Density or brightness

What does one does window width control?

Contrast

What is re-masking?

If the first mask image is not good enough you can choose a different one out of it

What is pixel shifting?

If the patient moves a little, you can shift the pixels over and lined them up better

What is view tracing?

Combining different parts of the vessel into one picture and seeing it fill up with contrast

Where is the x-ray tube located?

On the bottom

Where is the image intensifier located?

On the top

What does analog I I require?

CCD and ADC

Why must a patient couch/angiographic table be easy/convenient to use?

They must be comfortable because cases are long - 4 to 5 hours

What do pressure injector devices deliver?

A safe, uniform quantity of contrast under controlled conditions

What are the components of a contrast medium injection device?

Control panel, syringe, warming device, pressure mechanism

What does the control panel loud the tech to change?

Injection parameters

What is displayed on the control panel?

Flow rate, total volume, pressure (psi), rise rate, delay

What is flow rate?

Number of cc of contrast delivered every second, also the rate at which contrast is injected

What does the flow rate depend on?

The length of the catheter, internal diameter, number of side holes, injection pressure, viscosity

What is total volume?

Is the total of amount of contrast you're going to deliver

What is injection pressure (psi)?

Force needed for a specific dose of contrast medium

What is rise rate?

Time it takes to reach the desired flow rate, it's measured in seconds

What is delay?

Pause in injection rate/exposure

What is the warming device?

I thermal sleeve that maintains the temperature of the contrast material of syringe

What is the pressure mechanism?

This is what moves the plunger

What is quality control?

It includes an a valuation schedule of image processing/viewing systems, and internal x-ray beam, and ancillary equipment used

Why is traditional x-ray not great for vascular imaging?

It doesn't take pictures fasten off, it can't withstand heat, and it's not as flexible.

What are the three types of film for vascular imaging?

Rapid film changers, cut and roll film changers, cineangiographic cameras

What are other names for rapid changers?

Serial changers, or film changers

When was roll film changers developed?

As early as 1925

What parts are included on a roll film changer?

Grid, upper and lower intensifying screens, pressure plate, film cutter, supply and receiving magazines

What is the roll film changers max exposure?

12 per second

What is another name for cut film changers?

Puck film changer

What do you cut film changers allow?

Real-time imaging of angiograms

What is the main difference between roll film changers and cut film changers?

Cut film changers are precut films at the size of 14 x 14. Roll film changers are 14 x 17 that need to be cut.

What are the included parts of a cut some changer?

Grid, two intensifying screens, compression plate, compression table, supply magazines, receiving cassette

Why is it important to have good screen contact?

Because it will be blurry if it is not tight enough

What is the maximum exposure for cut film changers?

Six per second

What is the disadvantage of cut film changers?

They get jammed

What can rapid film changers be used in?

Either monoplane or biplane imaging

What is biplane?

When you have two planes, AP and lateral, taking images at the same time, with one contrast injection

What are the three components of cineangiography imaging systems?

Automatic brightness control, beam splitting mirror, cine camera filming device.

What is automatic brightness control?

It is used to compensate for KVP, mAs, or both, Giving you the best image possible, due to differences in anatomy.

What is the beams splitting mirror?

And allows you to view and record at the same time

What is cine camera filming device?

It takes many single images and move them fast enough, so it looks like one image. Usually at 15, 30, 60 per second. But when you have higher frame rate you have higher patient does.

What is synchronized system?

It means, the x-ray tube is energized only during the time when the cine film is in place for exposure

What needs to be included for digital image acquisition?

You need high frequency generator, x-ray tube, I I, image processor

When angio work is high-powered where shall it deliver?

Anywhere from 800-1000

Why are high frequency/high voltage generator is the most efficient?

Because they never drop to zero

What is the range of KV to better allow better visual of contrast in soft tissue?

50 to 100

What is it called when a generator switches on/off very rapidly to prevent you from overheating?

Pulse progressive fluoroscopy

What is interrogation time?

Time required to prepare x-ray tube to be switched on and reach the selected KVP and MA

What is extinction time?

The time required to stop exposure or shut off x-ray tube

What does three phase 12 poles or multi phase offer?

Constant potential 12 pulses per second and short exposure times at .001 ms

What is short term loading ability?

Amount of heat and anode can tolerate for a sequence of exposures/x-rays

What is continuous heat loading ability?

Amount of heat the anode can tolerate during a fluoro procedure

How big is a large diameter massive anode disc?

11 inches

What is the massive anode disc made of?

Tungsten, molybdenum, and rhenium

Maximum heat loading characteristics of an anode or measured in:

Heat units

What are the three x-ray tube rating charts?

Radiographic rating chart, Ano to cooling chart, housing cooling chart

What is the most important x-ray tube rating chart?

Radiographic rating chart

What do all radiographic tube charts plot?

MA, kVp, and time

What does each filament of each tube have?

A unique radiographic tube chart

What is the formula for heat unit?

HU = kVp x mA x time x rectification factor (x exposures)

What does
Single phase =
3 P 6 P =
3 P 12 P =
High frequency =

1
1.35
1.41
1.45

What does an anode cooling chart tell you?

How long it takes the anode to cool

What does a housing cooling chart tell you?

How long it takes to housing to cool, this is not used today

What does the heat loading capacity tube depend on?

Focal spot size, anode angle, anode speed, and generator

Why must you have a very small focal spot capability?

To view small vessels, to have more spatial resolution, so you get the best detail possible

What must you use if you have an analog image intensifier?

CCD and ADC

What does CCD replace?

TV camera tube

What is ADC?

Analog to digital converter

CCD connects to the ________ which connects________ then that_________ and that __________.

Output phosphor
Light to electrons
Signal is analog
Changes to CCD

What don't digital image intensifiers need?

Conversions, because they will go straight to the digital image

What is the last parts of digital image acquisition system?

Imaging processing hardware and computer

What are the modes of digital image acquisition?

Digital angiography
DSA
Digital Cineradiography

What does DSA stand for and what is it?

Digital Subtraction Angiography- process of removing or masking unnecessary anatomy, and Enhancing only anatomy of interest

What is digital angiography?

When you do one image at a time because it's used when only a few images is needed for placement of a PICC

What does PICC stand for?

Peripherally Inserted Central Catheter

What is mask/scout image?

The initial exposure before and contrast is injected

What is native image(s)?

Images with contrast

cineangiographic camera:

A specialized camera used for cardiac procedures to produce motion picture image of the flow of contrast medium through a vessel

What mm is used with cineangiographic camera?

35 mm roll is commonly used; 16 mm is also used, but not as much; higher dose with 35 mm

Cine Equipment has special tube rating charts and anode cooling charts; what are these known as?

Percent duty factor/cycle

What does window level control?

Density or brightness

What does one does window width control?

Contrast

What is re-masking?

If the first mask image is not good enough you can choose a different one out of it

What is pixel shifting?

If the patient moves a little, you can shift the pixels over and lined them up better

What is view tracing?

Combining different parts of the vessel into one picture and seeing it fill up with contrast

Where is the x-ray tube located?

On the bottom

Where is the image intensifier located?

On the top

What does analog I I require?

CCD and ADC

Why must a patient couch/angiographic table be easy/convenient to use?

They must be comfortable because cases are long - 4 to 5 hours

What do pressure injector devices deliver?

A safe, uniform quantity of contrast under controlled conditions

What are the components of a contrast medium injection device?

Control panel, syringe, warming device, pressure mechanism

What does the control panel loud the tech to change?

Injection parameters

What is displayed on the control panel?

Flow rate, total volume, pressure (psi), rise rate, delay

What is flow rate?

Number of cc of contrast delivered every second, also the rate at which contrast is injected

What does the flow rate depend on?

The length of the catheter, internal diameter, number of side holes, injection pressure, viscosity

What is total volume?

Is the total of amount of contrast you're going to deliver

What is injection pressure (psi)?

Force needed for a specific dose of contrast medium

What is rise rate?

Time it takes to reach the desired flow rate, it's measured in seconds

What is delay?

Pause in injection rate/exposure

What is the warming device?

I thermal sleeve that maintains the temperature of the contrast material of syringe

What is the pressure mechanism?

This is what moves the plunger

What is quality control?

It includes an a valuation schedule of image processing/viewing systems, and internal x-ray beam, and ancillary equipment used