RN Pacemakers

Pacers can be _________ or ___________

temporary or permanent

Asynchronous

generate impulses at fixed rate

Synchronous

generate a response to demand

Pacing can be

atrial, ventricular, atrio-ventricular or biventricular

Indications for pacing

conditions that interfere with heart intrinsic ability to initiate or conduct electrical impulses-- sinus node dysfunction, symptomatic bradyarrhythmias, tachyarrhythmias

Indications fro pacing 2

prophilaxis with post-cardiac surgery, biventricular pacing in HF to optimize CO-- EF </= 35%, IVCD >0.12 NYHA class III/IV

Temporary pacers

post open heart-- epicardial pacing

Transcutaneous or external pacing

temporary, use in emergent situations, stimuli delivered through defib system

Transvenous pacing

electrical stimuli delivered to endocardium surface of heart, delivered from external pulse generator or SC implanted generated

QRS deflection

impulses originate from endocardial surface look like LBBB; impulses originate from epicardial surgace look like RBBB

Firing

electrical impulse generated by pacer; a pacer spike

Capture

myocardial cells respond to electrical impulse

Sensing

detection of presence or absence of patient intrinisic electrical activity; capable in demand mode only

Rate control - fixed

asynchronous-- stimulus provided at preset rate, incapable of sensing intrinsic activity

Rate control- demand

synchronous-- stimulus provided when patient's rate falls below preset rate; must be able to adequately sense

Milliamps (mA)

amouont of current to elicit capture, mA

Pacing threshold

minimum current or mA needed for 100% capture

Sensitivity

allows pacer to detect intrinsic activity in demand mode; measured in millivolts mV-- lower number = greater sensitivity

Sense indictor

light flashes when inherent complex is detected

Threshold

minimum amplitude needed for detection by pulse generator; set sensitivity per policy; 1/2 of threshold

Failure to pace

generator off or faulty, battery dead

failure to capture

impulse from pacer, but not followed by appropriate wave form-- inadequate mA, electrode displaced, electrolyte imbalances--- pacer spike with no QRS

failure to sense (BAD BAD BAD)

pacer does not sense rhythm, sends sporadic spikes, can cause R on T--> v.fib. sensitivity TOO low (NUMBER IS HIGH), electrode dislodged, loose connections, depleted battery

Demand mode-- AAI

atrial paced and sensed, inhibited when P waves sensed

Demand mode-- VVI

ventricles paced and sensed, inhibited when QRS sensed, cannot restore atrial synchronicity

Demand mode-- DDD

designed to inhibit atrial and/or pacing if normal event is sensed

Undersensing

pacer fails to register depolarization of a chamber--

Oversensing

pacer detects signals other than depolarization of relevant chamber-- pacer can withhold pacing-- e.g artifact-- change sensitivity